LSRII #2

The LSRII (#2) is capable of detecting up to 14 parameters (Forward Scatter, Side Scatter, and eleven fluorescence detectors). It has three lasers for excitation of fluorochromes: a) a Coherent Sapphire laser for 488-nm excitation; b) a JDS Uniphase HeNe laser for 633-nm excitation; and c) a Coherent VioFlame for 405-nm excitation. In its basic configuration, the instrument is set up to detect 6 colors excited by the 488-nm laser, 3 colors excited by the 633-nm laser, and 3 colors excited by the 405-nm laser. This instrument uses BD Biosciences FacsDiva software for acquisition and analysis.
Basic Configuration
Detector |
Laser |
Wavelength |
Fluorochrome |
A |
488nm |
750-810nm |
PE Cy7 |
B |
488nm |
663-677nm |
PE Cy5/PerCP/7-AAD |
C |
488nm |
600-620nm |
PE Texas Red |
D |
488nm |
562-588nm |
PE/PI |
E |
488nm |
515-545nm |
FITC/GFP/Alexa Fluor 488 |
F |
488nm |
SSC |
------ |
A |
405nm |
600-620nm |
Qdot 605 |
B |
405nm |
564-606nm |
Qdot 585 |
C |
405nm |
425-475nm |
Pacific Blue/DAPI |
A |
640nm |
755-810nm |
APC Cy7/Alexa Fluor 750 |
B |
640nm |
690-735nm |
Alexa Fluor 700 |
C |
640nm |
650-670nm |
APC |
Scheduling
At this time, scheduling for analysis appointments on the LSRII must be scheduled through Research Core Facility personnel. It is recommended that you schedule your appointment one week in advance. All appointments are made on a first-come/first-served basis. Please understand that last minute scheduling cannot be guaranteed, and is generally unavailable.